Standard operating procedure for clean bench of medical purification workbench

This is a purification workbench that is widely used in the fields of medical and health, pharmaceutical, chemical experiments, etc., and provides a sterile, dust-free and clean environment.


Characteristics

1, vertical flow

2, using adjustable air volume fan system, light touch switch and stepless adjustment voltage. Ensure that the wind speed in the work area is always ideal.


Technical Parameters

model

SW-CJ-1F

Cleanliness

Level 100 @≥0.5μm (US Federal 209E)

Number of colonies

<0.5 pcs/dish. (diameter 90mm culture plate)

noise

≤62dB(A)

average wind speed

0.25-0.45m/s

Vibration half-peak

<=0.5um (x, y, z direction)

Illuminance

≥300LX

power supply

AC, single phase 220V/50HZ

Zui big power consumption

400W

weight

120KG

Working size

870*600*520mm

Dimensions

1050*680*1600mm

High efficiency filter specifications and quantities

865*555*38*1

Fluorescent / UV lamp specifications and quantities

16W*2/15W*2



model

SW-CJ-1FD

Cleanliness

Level 100 @≥0.5μm (US Federal 209E)

Number of colonies

<0.5 pcs/dish. (diameter 90mm culture plate)

noise

≤62dB(A)

average wind speed

0.25-0.45m/s

Vibration half-peak

<=0.5um (x, y, z direction)

Illuminance

≥300LX

power supply

AC, single phase 220V/50HZ

Zui big power consumption

400W

weight

120KG

Working size

870*600*520mm

Dimensions

1050*680*1600mm

High efficiency filter specifications and quantities

865*555*38*1

Fluorescent / UV lamp specifications and quantities

16W*1/15W*1


Overview: The SW-CJ-1F medical purification workbench is one of the most important equipments used for aseptic operations such as hygiene inspection and microbial culture.


Operating procedures:

1. Before use, it should be carefully checked, debug the various performances and conditions of the cleaning workbench, carefully check and debug the power supply, voltage is intact and stable, and use 75% alcohol to carefully clean the cleaning work before use.

2. Put the instrument supplies needed for the experiment into the workbench, turn on the main power supply, turn off the ventilation table and the general lighting switch, turn on the UV lamp for 30 minutes, and perform ultraviolet sterilization and disinfection. 3. Turn off the UV lamp, turn on the lighting, fan switch, adjust the wind to the appropriate size, wait for more than 30min to drain the ozone.

4. Wear a gown, wear a mask, and a hat. Hands and wrists are disinfected with 75% alcohol.

5. Ignite the alcohol lamp, avoid excessive items in the clean bench, use the straw, dropper, test tube, culture bottle, etc. before sterilizing.

6. Open the various types of caps before overheating to fix the dust; open the mouth of the bottle, the test tube mouth is over the flame, and the tweezers should be burned by flame before use.

7. The clean bench of the horizontal fan should be inclined, and the bottle mouth should be kept open and upright.

8. The same pipette or dropper should not be used continuously for several different cell lines; the pipette of the suction medium should be 0.5cm away from the culture bottle or the test tube mouth to avoid extending into the mouth of the culture bottle or the test tube; to prevent the cell line Mixed pollution.

9. The liquid leaking on the culture bottle or on the table is immediately wiped with an alcohol cotton ball.

10. After the experiment is completed, clean the work and keep the clean bench top clean and free of residue. Turn off the power after turning off the ventilation and lighting equipment. Take away all the rubbish when leaving the room.


Cleaning and maintenance procedures

1. After performing the aseptic experiment in the clean bench, the generated garbage should be treated separately according to the type. The discarded culture should be placed in a sterilization pot at 121 °C for 15 minutes, and the experimental products should be soaked in 5%. After 24 hours in the aqueous solution of Su, the rinse was taken out.

2. The net air filter is replaced once a year. If the blockage is serious, the replacement cycle can be shortened appropriately.


Shangyu Xingxing Instrument Equipment Co., Ltd. is located in Daoxu Town, Shangyu District, Shaoxing City, Zhejiang Province. It is an enterprise integrating production, product research and development, processing and trade. The company's main products: drying box series, incubator series, laboratory muffle furnace series, distilled water series, as well as clean workbench, air shower, fume hood, biological safety cabinet and other complete purification equipment.


The company has a strong technical research and development team, a strong after-sales guarantee service system, is the linger of the Chinese laboratory complete sets of equipment manufacturers, welcome customers to buy.



Imported Premium Wax

Polyethylene wax (PE wax), also known as polymer wax abbreviation polyethylene wax. It is widely used because of its excellent cold resistance, heat resistance, chemical resistance and wear resistance. In normal production, this part of wax as an additive can be directly added to the polyolefin processing, it can increase the gloss and processing properties of the product. As a lubricant, it has stable chemical properties and good electrical properties. Polyethylene wax is intermiscible with polyethylene, polypropylene, polyvinyl acetate, ethylene propylene rubber, butyl rubber. It can improve the fluidity of polyethylene, polypropylene, ABS and demultifying property of polymethacrylate and polycarbonate. Compared with other external lubricants, polyethylene wax has a stronger internal lubrication effect.

Imported Premium Wax,Imported Premium Ope Wax,Imported Premium Wax Emulsion,Imported Basf Wax Emulsion

Xingbang High Molecular Materials Co., Ltd. , https://www.chemicaladditive.com